background image

4-14

Chapter 4

Boonton Electronics

Remote Operation

4530 Series RF Power Meter

TKSDATA

Description:

Returns the sensor data table from sensor’s EEPROM. Note that year codes are excess-
1990, that is 5 means 1995, and 11 means 2001. Power and attenuation levels are coded as
dBm x 100; that is, -3500 means -35.00 dBm. All other values are in fundamental units.
Model numbers are only shown as the base number - any special “S” version number will
not appear.

Syntax:

TKSDATA?

Returns:

numeric array < type, model#, build month, day, year, serial#, calibration month, day, year,
attenuation, impedance, min pwr, max pwr, CW min pwr, CW max pwr >

Valid Modes:

All

Restrictions:

SYSTem:LANGuage

must be set to BOON to use native mode instructions

TKSSLOW

Description:

Returns the low bandwidth frequency calfactor table from sensor’s EEPROM. Frequencies
are in GHz and the calfactors are in dB. Count is the total number of data items in the string
including the upper and lower frequency limits.

Syntax:

TKSSLOW?

Returns:

numeric array < Count, LowerFreq, UpperFreq, Freq0, CF0, Freq1, CF1, Freq2, CF2... FreqN,
CFn >

Valid Modes:

All

Restrictions:

SYSTem:LANGuag

e must be set to BOON to use native mode instructions

TKSFAST

Description:

Returns the high bandwidth frequency calfactor table from sensor’s EEPROM. Frequen-
cies are in GHz and the calfactors are in dB. Count is the total number of data items in the
string including the upper and lower frequency limits.

Syntax:

TKSFAST?

Returns:

numeric array < Count, LowerFreq, UpperFreq, Freq0, CF0, Freq1, CF1, Freq2, CF2... FreqN,
CFn >

Valid Modes:

All

Restrictions:

SYSTem:LANGuage

must be set to BOON to use native mode instructions