Remote Operation

4.9 ERROR AND STATUS CODES.

The following table shows codes that will be returned in response to any command that is unrecognized, invalid, or
not applicable in the current state. A list of these errors may be read in the sequence they occurred using the SCPI
command

SYSTem:ERRor

. This command will report each error by code and a short, quoted text description. An

error code of zero corresponds to “No Error”.

Table 4-2. Remote Interface Error Codes

E-100

User sent a command which was not recognized.

E-101

User sent a command that included a sub-command field, where none was expected.

E-103

User sent a command that did not include a sub-command field, where one was expected.

E-108

User included arguments, however none were expected.

E-109

Argument expected, but not found.

E-121

Arguments were present, but wrong type (such as alpha-text present where only numerical data
is expected).

E-140

Remote command string from host overflowed instrument’s input buffer.

E-200

Cannot select channel 2 on a single channel instrument.

E-213

Cannot set INITiate:IMMediate (start a single measurement) while INITiate:CONTinuous is set
to ON (free-run mode).

E-214

Cannot perform read command while trigger source is set to bus.

E-221

Attempting to set a parameter or talkmode while in the wrong measurement mode (such as
setting a statistical talkmode while in peak-continuous mode, or setting the marker in percent
while in peak-pulse mode, or using a Boonton language command in an SCPI-only mode).

E-222

Numeric argument out of range.

E-224

Text argument does not match a valid list entry.

E-230

Fetch command performed and no valid data available.

E-340

Error during sensor calibration. Check connections.

Table 4-3. Measurement Result Status Codes

-1

Measurement is STOPPED. Value returned may be “stale data”.

0

Error return. Measurement is not valid.

1

Normal return - no error. Measurement is valid.

2

An Over-range or Under-range condition exists.