Sper Scientific 850038 Pure Water Meter User Manual

Page 11

Advertising
background image

K=0.9

66

When should you do the calibration?

To completely calibrate the meter,
clear all

previous calibration data,

(t

o erase all calibration data,

see P7.1 on page 32 (Fig. G)).

To

normalize to another value, see

P4.2 on page 28

(Fig. F). Before

the calibration

standard value of the normalized temp.
must be known. (Refer to the datasheet
enclosed with
your solution.)

resetting this value,

If the conductivity of measured solutions
are < 100 S, or TDS are < 50 ppm,
calibrate the meter at least once
a week for accuracy.

o

C

Fig .F

Fig .G (8306)

For first us

eand for best results,

use solution to calibrate or
manually input the cell constant

(refer to

the label on the probe).

NOTE: The cell constant has
been programmed into meter.

if

solution is not

readily available.

Cell Constant

If the meter is used in the mid ranges,
it is needed to do calibration at least
once a month.

If the measurement is performed at
extreme temperatures, calibrate at
least once a week.

10

K=0.966