c)P2.0: Meter configuration: (CoF)

P2.2:

READY indicator:(rdy)

Fig. Q

Fig. R

25

P2.1: TDS factor: (tdS)
The concentration of dissolved salts
in solution increases the conductivity.
This effect varies from salt to salt and
is roughly linear in a given range for
a given salt. The TDS conversion
factor is a value used by the meter
to convert from conductivity to TDS.

After selecting P2.0,

press ENTER

to select P2.1. Press again to enter
P2.1. TDS factor flashes on the LCD
(Fig. Q). You can press UP/DOWN
to change the value from 0.40 to 1.00.
The default value is 0.50.Press
ENTER to confirm
the TDS factor and
select P2.2
automatically.

Turn the

READY indicator

"on" to alert you that the measurement
is stable.
Select

"off" for faster response.

Press UP/DOWN to switch the
ready function to "on" or "off". (Fig. R)
Press ENTER to confirm the last state
and select P2.3 automatically.