8.Press the UP/DOWN keys to match
the value on the primary display to
the value of the calibration standard.

Adjust the SALT reading up
to +20% from the detected value. If
the detected value and standard
values differs by more than +20%,
clean or replace probe

The maximum input value is 50 ppt
and the minimum input value is 1ppt.

9. When "Ready displays, press
"ENTER" to confirm the calibration.
The meter switches to SALT

measurement mode. Note:For high
accuracy in

Conductivity &TDS

modes, perform the calibration
again before measuring Conductivity
or TDS if ou have performed a
salt calibration.