PREPARATION FOR CALIBRATION

Two issues should be considered
before operation:
W

hat is the right calibration standard?

W

hen should you calibrate?

Selecting a calibration standard

For best results, select a conductivity,
TDS or NaCl

standard near the

s

ample value you are measuring.

Alternatively, use a calibration solution

value which is approximately 2/3 of the

full scale of the measurement range you

plan to use.

For example, in the 0 to 1999 uS range,
use 1413 uS solution for calibration.

Conductivity

measuring range

Recommended cal.

solution range

TDS measuring

range (factor=0.5)

Recommended cal.

solution range

0~19.99uS

0~199.9uS

0~1999uS

0~19.99mS

0~199.9mS

6.00~17.00uS

60.0~170.0uS

600~1700uS
6.00~17.00mS

60.0~170.0mS

0.00~9.99ppm

0.0~99.9ppm

0~999ppm

0.00~9.99ppt
0.0~199.9ppt

3.00~8.50ppm
30.0~85.0ppm

300~850ppm
3.00~8.50ppt
30.0~85.0ppt

1

2

3
4

5

1

2

3
4

5

R

efer to below table.

Use the recommended solution for
different conductivity and

TDS ranges

for best results.

DO NOT reuse the calibration solution.
Contaminants in the solution will affect
the calibration and the accuracy. Be
sure to use fresh solution each time.