Sper Scientific 850038 Pure Water Meter User Manual

Page 32

Advertising
background image

If constant K=0.1/1/10 does not
completely meet your needs,
input the cell constant after selecting
the K=0.1, 1 or 10.
For example: K=0.992
After saving P5.1, P5.2 will be
selected. Press ENTER again to
advance to

P5.2.The cell constant r

(0.1, 1 or 10)

will flash on the LCD.

Press UP to select the flashing value
from 0~9. ( Fig. AA)
Press DOWN to change the flashing
digit.
The selectable range is ell
constant which is selected in P5.1.
Press ENTER to confirm the last
value and return to P5.0.
NOTE:

20%of c

After inputting cell constant,

all calibration information in P8.0 will
be cleared. Manual input of the Cell
Constant

will change all the constants

in ranges

1~5 at the same time.

K=

g).P6.0: Memory Clear: (Clr)

Fig. AA

31

P5.2: Input the cell constant:(InPt)

Use this function to clear all stored
memory when you need to store a
new series of values. This function
lets you avoid confusing the old
values with the new ones.
In P6.0, p

ress ENTER to enter P6.1.

Press UP/DOWN to select "n"-NO or
"y"-YES, then press ENTER to
confirm and the meter will return to
P6.0. (Fig. AB)

P6.1: Memory clear(CLr)

Advertising