Sper Scientific 850038 Pure Water Meter User Manual
Page 33

h)P7.0:Reset to factory default setting(rSt)
P7.1: Reset (rSt) restores the factory
i)P8.0:View calibration data (CAL)
If the memory is full, you must clear
all values to continue recording. If
not, when you record the 100th value,
"FuL" will flash 3 times.
.
Fig. AB
Fig. AC
32
default settings. This function clears
all calibration data and other user
defined parameters except the clock
settings and memory.
In P7.0, p
ress ENTER to advance to
P7.1.
Press UP/DOWN to select "n"-NOr
or
"y"-YES. (Fig. AC)
Press ENTER to confirm and then
return to P7.0.
NOTE: Refer to
he default parameters or recalibrate.
Appendix A to review
t
When replacing a probe, clear all
calibration data in memory.
This function lets you recall previous
calibration data and may help you
know when its time to re-calibrate
the meter. This function is for
"Review" purpose only.