Adjust the normalization temperature, The m eter default is 25 c (77, Fig x – Sper Scientific 850038 Pure Water Meter User Manual
Page 29
A
djust the normalization temperature
o
o
from 15 to 30 C (59 to 86 F).
o
o
The m
eter default is 25 C (77
F).
After pressing ENTER, the normali-
zation temperature will flash on the
LCD. Press UP/DOWN to change the
o
value from 15.0 to 30.0 C (59.0~
o
86.0
F). (Fig. X)
Press ENTER to confirm the last
value and advance to P4.3.
Select P4.0 and press ENTER to
advance to
P4.1. Press ENTER again
and the
Temperature Coefficient will
flash
. Press UP/DOWN to change
the value from 0.0 to 10.0, the unit is
o
%/ C (Fig. W). Press ENTER to
confirm the last value and advance to
P4.2 automatically.
The meter will normalize its cond.
measurement to a standard temp.
which you preset.
Fig W
o
C
Fig X
o
C
NOTE: To know more about the
temperature effect on on measurement,
refer to Appendix D on page 43
28
P4.2: Normalization temperature:(nor)