TDS MEASUREMENT

To measure with manual temperature
compensation, follow the below
steps:
1. Turn on the meter. Press the set key

R

efer to P2.4 (page 26)

to disable the ATC function.

2. To set a manual temperature
compensation value. See

P4.3 on page 29.

3. To switch the meter back to normal
mode. The middle region of LCD will
display a fixed temperature that you
input, "ATC" is not displayed.
To start the measurement
follow steps 4-6 on page 19.

more than one second to enter setting
mode.

Manual Temperature Compensation
(

For COND. & TDS)

NOTE:
Manual Temperature Compensation
only available when the ATC
is off. For non-compensated
measurements, change the temp.
coefficient to 0.0%.( Fig. K)

o

C

Fig. K

1. Turn on the meter.
2. Set the TDS conversion factor to a
correct value. The factory default
value is 0.50. Refer to P2.1 on page
25.

Fo

llow the below steps for TDS

measurement.