Sper Scientific 850038 Pure Water Meter User Manual
Page 25

uS
The LCD will display E03 if the
measured conductivity/TDS values
are beyond the limit value.
S
elect a range within limits.
The meter will reset to the Auto-
ranging function when it is turned off.
The
manual ranging function must be
set
each time the meter is turned on.
Type
P1.1
P1.2
P1.3
P1.4
P1.5
Range
Ist Range
2nd Range
3rd Range
4th Range
5th Range
Cell Constant
=0.1
Cell Constant
=1.0
Cell Constant
=10
0~1.99uS/ppm
0~19.99uS/ppm
0~199.9uS/ppm
0~1999uS/ppm
0~19.99mS/ppt
0~19.99uS/ppm
0~199.9uS/ppm
0~1999uS/ppm
0~19.99mS/ppt
0~199.9mS/ppt
0~199.9uS/ppm
0~1999uS/ppm
0~19.99mS/ppt
0~199.9mS/ppt
0~1999mS/ppt
The digits flash on the display.
Fig. P
CON
24
Press the UP or DOWN key to select
from P1.1 to P1.5 and then press
ENTER to save.
NOTE: Press ENTER to select
P1.1or other ranges, LCD will flash
the limit value (Fig. P). Press the
ENTER
key to confirm. The meter will
return to normal measure mode.