Performing an immediate data transfer – Vaisala viewLinc 4.3 User Manual

Page 155

Advertising
background image

Transferring Data

viewLinc 4.3

4

|

4

147

6

In the Destination area, complete the fields:

Destination

: Choose the destination directory for the saved

data (.spl) file (do not use a mapped drive path).

Sample directory paths could include: c:\Device_files or
\\<machinename>\<destination_dir>. We recommend you
use a directory on the viewLinc Host Server (for example,
c:\<destination_dir>).

Note: During a data transfer, you may be prompted to enter a

valid username and password to save the file to a target dir-
ectory or shared directory. You also need to ensure that the
account which installed viewLinc (that is, which runs the
viewLinc Enterprise Server service) has sufficient per-
missions to write to the directory where you want the trans-
ferred data to go. For more on Services with viewLinc, see
“FAQs & TROUBLESHOOTING” on page 161.

Field 1/2/3

: Specify the order you want to used to identify the

three fields which will make up the transfer filename. File-
names for .spl files are created from combinations of the
Device's description, serial number, and the date and time
the file was saved.

For example, Device ID-08094523-2011-04-22 16-30-01.spl.

7

To test that the transfer will complete as defined, click Transfer
Now.

Use Windows Explorer to ensure the data file transferred to

the specified file location.

8

Click Save to save these settings for a scheduled transfer.

Note: If comments are enabled from system preferences, the Add

Comment window appears. Select a preconfigured comment or
enter your own. A row containing information on the sched-
uled transfer for each Device you’ve set up appears in the
Event Log.

Performing an Immediate Data Transfer

You may be required to initiate an immediate transfer, before a
scheduled transfer is due, perhaps after receiving an alarm that Logger
memory is running low.