3 step 3 – loggernet set-up, Step 3 – loggernet set-up, The connection type – Campbell Scientific RF401-series and RF430-series Spread Spectrum Data Radios/Modems User Manual

Page 22: 2. lacking 12 v on cs i/o pin 8

Advertising
background image

RF401-series and RF430-series Spread Spectrum Radio/Modems

TABLE 3-2. Lacking 12 V on CS I/O Pin 8

EQUIPMENT

SERIAL NUMBER

CR500

< 1765

CR7 700X Bd.

< 2779

21X

< 13443

CR10 Wiring Panels

All (black, gray, silver)

PS512M Power Supply

< 1712

When you connect power to the RF401 (through the SC12 cable or the

optional Field Power Cable) you should see the power-up sequence of red

and green LEDs described in Step 1 (assuming datalogger is powered).

Current dataloggers and wiring panels (not mentioned in TABLE 3-2)

provide 12 V on pin 8. For older products not listed, check for 12 V

between CS I/O connector pin 8 and pin 2 (GND) or contact Campbell

Scientific.

c. Configure the remote radio using the same settings as the base radio (see

“e” in Section 3.2.1, Step 1 – Set up Base RF401 or RF430). This

assumes the base radio is set to AutoSense.

3.2.3 Step 3 – LoggerNet Set-up

a. The next step is to run LoggerNet and configure it to connect to the

datalogger via the radio’s point-to-point or PakBus network you have set

up. For point-to-point and PakBus communications, the radios are

represented as direct connect in LoggerNet (see FIGURE 3-2). Refer to

Section 5.2, LoggerNet, for the LoggerNet setup of point-to-multipoint.

FIGURE 3-2. Point-to-Point or PakBus communications, select direct

connect as the connection type.

10

Advertising