Myron L 729II User Manual

Page 17

Advertising
background image

d. Install 5.49kΩ resistor at RED (RD) and the

NEUTRAL (NU) connector locations, as shown in

figure II.E.5.

5. Carefully reinstall the front panel, bottom first. Ensure

no wires have been pinched between enclosure and

front panel.

6. Reinstall the four (4) screws and tighten.

7. To operate, turn power ON.

NOTE: Recalibration will require both the solution and sensor

be at 25°C for maximum accuracy.

4. SOLID STATE OUTPUT

24 VDC Unregulated 30mA max. The following instructions are

assuming the Monitor/controller enclosure is already open.

a. Piezo Electric Alarm - PA/PAO (option)

For additional information, see Piezo Alarm under Options in

section III.I.

1. If not already installed, peel tape backing from PIEZO

and press into place per figure III.I.3.

2. Attach connector to main control circuit board per

figure III.I.4.

NOTE: If remotely mounted; cut wires and splice as necessary,

use comparable wire. Piezo requires 1/4” (6.35mm) hole in user

panel.

b. Remote Alarm - RA™ (option)

For additional information, see RA Instructions under options in

section III.J.

1. Run user supplied #22, 2 conductor speaker type wire

from Monitor/controller to RA location as necessary.

Additional wire may be ordered, part #RAW-200, see

Options & Accessories.

2. Open the RA by removing the four (4) screws.

3. Locate and remove the 8” 2 conductor wire attached

to RA.

4. At the controller, connect the extension wires to the 8”

2 conductor wire with the wire nuts provided — Black

to Positive (+) and White to Negative (-). Be sure to

first pass the wire through the user supplied waterproof

strain relief in the enclosure.

5. Plug the reddish brown female connector into the

male connector on the controller CB marked either RA

or PA (see inside case label for location). It will only go

on the connector one way.

6. At the RA, connect the wires to the connector — Black

to Positive (+) and White to Negative (-).

7. To test, simply turn on the controller and adjust

controller set point until the alarm/piezo sounds off. If

controller is not yet connected to a functioning sensor,

on conductivity/TDS controllers it will be necessary to

press and hold the Full Scale test switch.

The black button on the front of the RA will mute the

piezo alarm for approximately three minutes or until

you improve the water quality (readjust controller set

point). The piezo alarm will continue to sound off every

three minutes until the user has improved the alarm

condition inside the controller. If three minutes muting

is fine for your application, skip to step 9.

8. If three minutes is too long or too short, adjust time

delay control inside RA until desired mute time is achieved

(adjustable from approximately 6 seconds to

10 minutes).

9. Replace the bottom of the RA, and secure RA to the

surface you have selected for its installation.

NOTE: If the RA does not sound off;

1. Check the polarity of the extension wire connections.

2. Be sure the controller is actually switching (relay will

click).

13

BK WT RD GN NU R- R+

FS SW

3S

SENSOR LEADS

INSTALL 10KΩ

RESISTOR HERE

CAL

REMOVE THESE

THREE LEADS

FS SW

BK WT RD GN NU R- R+

3S

SENSOR LEADS

CAL

SENSOR LEADS

FS SW

3S

CAL

INSTALL 100KΩ

RESISTOR HERE

INSTALL 5.49KΩ

RESISTOR HERE

BK WT RD GN NU R- R+

Conductivity/TDS Main CB Assembly

Figure II.E.3

Resistivity Main CB Assembly

Figure II.E.4

Resistivity Main CB Assembly

Figure II.E.5