Myron L 729II User Manual

Page 50

Advertising
background image

46

Sensor Connections

For convenience, it is recommended you connect the Primary or

controlling sensor to SENSOR 1 position as shown in

figure III.H.1.

CAUTION: The sensor connectors require only a small screw-

driver or a pen to push on the release lever. Follow the color

code as marked. The release levers may be broken or damaged

if not pushed straight toward the CB. DO NOT push the release

levers sideways.

3. ALARM/CONTROL CONFIGURATION

Sensor Switch

As shipped from the factory, the 3SR Module is configured so

that only the Primary or First switch position “S1” is set to alarm/

control, figure III.H.2. If your particular application requirements

are different, the following will help you to reconfigure the alarm/

control function to another sensor or add a controlling sensor.

Pick the appropriate scenario below.

NOTE: the setpoints will not change nor are they separately

adjustable for each sensor.

If adding an alarm/control sensor.

1. Locate the 8 position select switch on the 3SR Module.

2. Set appropriate select switch UP or ON, i.e. to add S3

sensor, set switch marked S3 UP or ON, see

figure III.H.3.

If changing alarm/control from one sensor to another, i.e. S1 to

S2 sensor.

1. Locate the 8 position select switch on the 3SR Module.

2. Set the S1 select switch in the DOWN or OFF position.

3. Set the S2 select switch in the UP or ON position, see

figure III.H.4.

To disable the Sensor alarm/control function completely.

1. Locate the 8 position select switch on the 3SR Module.

2. Set all X prefixed and S prefixed select switches in the

DOWN or OFF position, see figure III.H.5.

Range Switch

As shipped from the factory, the 3SR Module is configured so

that only the Primary or First switch position X1 is set to alarm/

control, figure III.H.6. If your particular application requirements

are different, the following will help you to reconfigure the alarm/

control function to another range or add a controlling range. Pick

the appropriate scenario below. The setpoints will not change

nor are they separately adjustable for each range.

NOTE: The Range Switch is labeled:

SENSOR

SWITCH

BACK/CONNECTOR SIDE

-121

2000µS

UP

}

3SR MODULE

SENSOR OUTPUT

TO SENSOR INPUT

SENSOR INPUT

3S

MAIN CIRCUIT BOARD

RANGE

SWITCH

Figure III.H.1

SENSOR 1

SENSOR 2

SENSOR 3

SENSOR 1

BK WT RD GN NU

BK WT RD GN NU

BK WT RD GN NU

BK WT RD GN NU R- R+

SENSOR 2

SENSOR 3

3S

ON

SELECT SWITCH

Figure III.H.2

ON

Figure III.H.3

ON

Figure III.H.4

Figure III.H.5

ON

ON

Factory Default

3 SENSOR/3 RANGE INPUT MODULE 3SR

NOT INSTALLED ON

MONITOR ONLY MODELS.

COND RES X1 X10 X100 S1 S2 S3

COND RES X1 X10 X100 S1 S2 S3

COND RES X1 X10 X100 S1 S2 S3

COND RES X1 X10 X100 S1 S2 S3

COND RES X1 X10 X100 S1 S2 S3

Advertising