42

figure III.F.4.

To disable the Sensor alarm/control function completely.

1. Locate the 8 position select switch on the 3S Module.

2. Set all S prefixed select switches in the OFF position,

see figure III.F.5.

3SE MODULE with

ENCLOSURE

If adding an alarm/control sensor.

1. Locate the 8 position select switch on the 3S Module.

2. Set appropriate select switch DOWN or ON, i.e. to add

S3 sensor, set switch marked S3 ON, see figure III.F.8.

If changing alarm/control from one sensor to another, i.e. S1 to

S2 sensor.

1. Locate the 8 position select switch on the 3S Module.

2. Set the S1 select switch in the UP or OFF position.

3. Set the S2 select switch in the DOWN or ON position,

see figure III.F.9.

To disable the Sensor alarm/control function completely.

1. Locate the 8 position select switch on the 3S Module.

2. Set all S prefixed select switches in the UP or OFF

position, see figure III.F.10.

3SE REASSEMBLY

1. Carefully reinstall the front panel, bottom first. Ensure

no wires have been pinched between enclosure and

front panel.

2. Reinstall the four (4) screws and tighten.

3. To operate, turn power ON.