c. Hysteresis (Dead Band) Adjustment*

The hysteresis or dead band is approximately ±3% of the set

point at full scale as it leaves the factory. Under normal (most)

conditions it will not be necessary to adjust.

However, if you desire to make an adjustment please keep the

following in mind.

The adjustment is very simple and is based on set point loca-

tion. If the set point* is in the upper 75-100% of the scale, the

hysteresis control pot should be turned fully to the right.

If the set point is in the lower portion of the scale, i.e. 5-25% of

scale, the control pot may be turned fully to the left.

If you are operating in the center, 25-75% of scale, the control

pot may be adjusted in the center.

Or the hysteresis or dead band may be adjusted to tighten the

control of a particular process.

CAUTION: Adjusting the hysteresis too narrow may cause the

alarm to fluctuate (on-off) due to flow, chemical mixing or bub-

bles causing the relay to chatter. This condition is to be avoided,

it could damage your valves, pumps, etc. and will eventually

damage the relay.

* Applies to both set points.

The following is assuming the front panel is already removed and

the set points have been set, if not, see “Set Point Adjustment”.

1. Turn power ON.

2. Locate the Hysteresis Control (HYS1) located next to

the Set Point #1 (SP1) adjustment - it is a single turn

pot. See figure III.A.4.

3. Adjust as described in “Hysteresis (Dead Band)

Adjustment” or as desired.

4. Press set point test switch on the front panel to verify

Hysteresis is set as desired.

5. Repeat for SCO Module. See figure III.A.4.

6. Press set point test switch on the front panel to verify

Hysteresis is set as desired.

7. Turn power OFF.

d. Second Relay Connection

1. Place the user supplied relay interface cable and

watertight cable restraint into the enclosure’s

appropriate access hole.

2. Neatly connect wires to the relay connector as shown

in figure III.A.4.

e. Solid State Output Options

24VDC Unregulated 30mA max.

Same as 4. above, SOLID STATE OUTPUT.

1. Piezo Electric Alarm - PA (option)

Plug PA connector to CB as shown (See

Figure III.A.4).

2. Remote Alarm - RA™ (option)

Plug RA connector to CB as shown (See

Figure III.A.4).

3. Connect to your own alarm or ?

Ensure your requirements do not exceed the 24VDC

Unregulated 30mA maximum.

Ensure the polarity is correct.

Attach your wires.

Attach harness connector to controller connector (RA)

per figure III.A.4.

If necessary, you may order connector with 8” Harness

part #RAH, from the Myron L Company.

REASSEMBLY:

1. Carefully reinstall the front panel, bottom first. Ensure

no wires have been pinched between enclosure and

front panel.

2. Reinstall the four (4) screws and tighten.

3. To operate, turn power ON.

Figure 5

FRONT PANEL

Rear View

Primary/Main CB

Red LED

Primary/Main

CB Set Point

Switch

SC/SCO

Red LED

SC/SCO

Set Point

Switch

Green

LED

Figure 6

HIGH

LOW

SET POINT

750II

MICROSIEMENS / CM

FRONT PANEL

Primary or Main CB

Alarm/control Location

SC/SCO

Location

22

Advertising
This manual is related to the following products: