Myron L 729II User Manual

Page 45

Advertising
background image

Module to the main Monitor/controller CB at the sensor

input location as shown on label.

5. Attach the separate small cable harness from the 3 Sensor

Input Module to the small 2 pin male connector located

to the left of the sensor input connectors on the main

CB. NOT INSTALLED ON MONITOR ONLY MODELS.

Sensor Connections

For convenience, it is recommended you connect the Primary or

controlling sensor to SENSOR 1 position. See figures 1 or 6 for

3SE.

CAUTION: The sensor input connectors require only a small

screwdriver or a pen to push on the release levers. Follow the

color code as marked. The release levers may be broken or

damaged if not pushed straight toward the CB. DO NOT push

the release levers sideways.

3. ALARM/CONTROL CONFIGURATION

As shipped from the factory, the 3S Module is configured so that

only the Primary or First switch position SENSOR 1 is set to

alarm/control as shown in figure 2, or figure 7 for 3SE. If your

particular application requirements are different, the following

will help you to reconfigure the alarm/control function to another

sensor or add a controlling sensor. Pick the appropriate scenario

below.

NOTE: the setpoints will not change nor are they separately

adjustable for each sensor.

3SO MODULE

If adding an alarm/control sensor.

1. Locate the 8 position select switch on the 3S Module.

2. Set appropriate select switch UP or ON, i.e. to add S3

sensor, set switch marked S3 ON, see figure III.F.3.

If changing alarm/control from one sensor to another, i.e. S1 to

S2 sensor.

1. Locate the 8 position select switch on the 3S Module.

2. Set the S1 select switch in the DOWN or OFF position.

3. Set the S2 select switch in the UP or ON position, see

41

Figure III.F.4

ON

Factory Default

Figure III.F.2

ON

Figure III.F.3

ON

Figure III.F.5

ON

COND RES X1 X10 X100 S1 S2 S3

COND RES X1 X10 X100 S1 S2 S3

COND RES X1 X10 X100 S1 S2 S3

COND RES X1 X10 X100 S1 S2 S3

SENSOR INPUT

3 S

MAIN CIRCUIT BOARD

SENSOR

SWITCH

SENSOR 1

SENSOR 2

SENSOR 3

3SR MODULE

3S

COND

RES

X1

X10

X100

S1

S2

S3

SENSOR

OUTPUT

SENSOR 1

SENSOR 2

SENSOR 3

ON

SELECT

SWITCH

COND

RES

X1

X10

X100

S1

S2

S3

BK WT RD GN NU R- R+

3

S

NOT INSTALLED ON MONITOR ONLY MODELS.

SENSOR 3

SENSOR 2

SENSOR 1

BK WT RD GN NU

BK WT RD GN NU

BK WT RD GN NU

SELECT

SWITCH

ON

3 SR MODULE

SENSOR

OUTPUT

BACK/CONNECTOR SIDE

3 SENSOR INPUT MODULE (3SE)

Figure III.F.6

ON

Figure III.F.8

ON

Figure III.F.9

ON

Figure III.F.10

ON

Figure III.F.7

Factory default

COND

RES

X1

X10

X100

S1

S2

S3

COND

RES

X1

X10

X100

S1

S2

S3

COND

RES

X1

X10

X100

S1

S2

S3

COND

RES

X1

X10

X100

S1

S2

S3

Advertising