31

counterclockwise position. If not, turn SCO Set Point

control fully CCW - it may be 30 turns - or until it clicks.

For location see figure III.C.13.

3. Turn power ON.

4. Press BOTH the Temperature Select switch and the

Set Point switch on the front panel.

5. Adjust the TP/TPO Module Set Point Two (SP2) to

temperature as desired. See figure III.C.1.

6. Turn power OFF.

c. Hysteresis is fixed at approximately ±6°C unless

Special Ordered with adjustable control.

Hysteresis with adjustable control is variable from ±0.6 to 6°C,

see figure III.C.1 for location.

CAUTION: Adjusting the hysteresis too narrow may cause the

alarm to fluctuate (on-off) due to temperature variations causing

the relay to chatter. This condition is to be avoided, it may dam-

age the heater, chiller, etc. and will eventually damage the relay.

R

EASSEMBLY

1. Carefully reinstall the front panel, bottom first. Ensure

no wires have been pinched between enclosure and

front panel.

2. Reinstall the four (4) screws and tighten.

3. To operate, turn power ON.

Figure III.C.12

FS

INC

INC

TEMP

CONTROL

HARNESS

DEC DEC

SET POINT

CONVERSION (SPC)

SPC

Figure III.C.13

SC/SCO MODULE

(SECOND ALARM/CONTROL)

MAIN CIRCUIT BOARD NOT SHOWN FOR CLARITY

SCO SET POINT #2 ADJUST

(TURN CONTROL FULL CCW)

SPC