I. PIEZO ELECTRIC ALARM

-PA

Piezo Electric Alarm ordered with Digital

Monitor/controller.

PAO Piezo Electric Alarm ordered separately for Digital or

remotely mounted Analog Monitor/controller or OEM

models.

1. DESCRIPTION

The PA/PAO is an electronic sound device capable of emitting a

80dB or more @ 30cm, high pitched squeal. See figure III.I.1.

Specifications

Oscillating Frequency — 3.0≠0.5KHz

Operating Voltage (750II) — 24VDC Nom. (1.5-30VDC Max.)

Sound Pressure Level (Min) 30cm/12VDC — 80dB

Current consumption (Max) @ 12VDC — 12mA

Tone — Constant

Operating Temperature — -20 - +60°C

Size — 24 x 9.5 mm

2. INSTALLATION

Briefly -

The PA/PAO Piezo Alarm attaches to the front panel with the

tape supplied, see figure III.I.3.

The wire harness plugs into a methode connector on the main

CB*, see figure III.I.4.

* May also be connected to the SC Module as shown in figure

III.I.5, if desired or so equipped.

WARNING: BEFORE STARTING, IF MONITOR/

CONTROLLER IS INSTALLED, ENSURE THE POWER IS OFF.

FAILURE TO DO SO COULD CAUSE DAMAGE TO THE IN-

STRUMENT, AND COULD BE HARMFUL OR FATAL TO PER-

SONNEL. ONLY QUALIFIED PERSONNEL SHOULD INSTALL

OR SERVICE ELECTRICAL EQUIPMENT.

Physical

NOTE: Remote mounting will require a .25” (6.35mm) hole, and

extending the harness. Use #22 gauge speaker wire. Observe

polarity.

If the front panel has all ready been removed from the enclosure

skip to #4.

1. Using a standard slot screwdriver remove the four (4)

screws on the front panel.

2. Carefully wiggle the front panel to loosen.

Pull gently toward you. Do not pull more

than about 8 inches/20CM or you could damage the

wiring harness.

3. Turn the front panel around so that the back side is

facing you.

4. Using a 1/4” drill or a sharp knife, carefully open the

CENTER front hole in the front panel as shown in

figure III.I.2.

5. Peal off tape backing from PA, and install between

guides as shown in figure III.I.3.

6. Set front panel aside.

7. Connect wire harness to Main CB as shown in

figure III.I.4, or to SC as shown in figure III.I.5.

Test

1. Turn power ON. Depending on configuration, the Piezo

will sound off when —.

NOTE: If the sensor is connected, the solution value, set point

value and Set Point Conversion (SPC) jumpers may affect the

test.

a. For a Conductivity/TDS Monitor/controller, as

shipped from the Myron L Company, the alarm will

!

CAUTION - READ FOLLOWING CAREFULLY

Figure III.I.1

Double Backed Tape

Harness ~12"

Piezo Alarm

Methode Connector

Figure III.I.2

FRONT PANEL

Rear View

CUT OPEN

THIS HOLE