43

G.

3 RANGE SWITCH MODULE

3RO

3 Range Switch Module with 3 foot / 1 meter cable

ordered separately (customer installed and

mounted).

3RE 3 Range Switch Module with 3 foot / 1 meter cable

AND separate Enclosure (as an accessory).

Enclosure is the same size as the

Monitor/controller - 4.8”W x 6” H x 3.9”D.

1. DESCRIPTION

The 3 Range Switch Module (3RO or 3RE) is a circuit board

(CB) with a 3 position rotary switch. This option allows the user

to select which range to view, and to set for alarm/control func-

tion. The 3R may be configured - via select switches - to disable

or stop the alarm/control function from operating when chang-

ing ranges. The 3R will function with all Conductivity/TDS and

Resistivity models, analog or digital.

NOTE: the setpoint(s) are NOT separately adjustable for each

range.

It is available in multipliers of “10” only.

Example #1 - Conductivity/TDS: 758II-117-3R = 500µS (X1)

w/ 5000µS (X10) & (X100) 50mS additional ranges.

Example #2 - Resistivity: 753II-11-3R = 20MΩ (X1) w/ 2MΩ

(X.1) & 200KΩ (X.01) additional ranges.

As shipped from the factory, the number “1” Range position is

the primary or the alarm/control position while the other posi-

tions are monitoring ONLY. In other words, while in the #2 (X.1)

or #3 (X.01) position the alarm/controller is disabled. In the

above example #2 the 20MΩ (X1) is the primary or alarm/control

range.

Specifications

NONE

2. INSTALLATION

Briefly -

The 3RO must be installed/mounted by user outside of the

Monitor/controller housing by drilling a hole in user panel. For

the 3RE, a cutout the size of the enclosure is required.

On a Conductivity Monitor/controller the Range Module is

removed from Monitor/controller CB and re-installed on the 3R

Switch Module. On a Resistivity Monitor/controller the Range

Module is removed from Monitor/controller CB and a special 3

Range-Range Module is installed on the 3R Switch Module.

A cable harness from the 3R Module is connected to the main

Monitor/controller circuit board in place of the Range Module.

A separate small connector & harness is for the alarm/control

disable or kill, and attaches to the small 2 pin male connector lo-

cated to the left of the sensor input connectors on the main CB.

NOT INSTALLED ON MONITOR ONLY MODELS.

WARNING:

BEFORE

STARTING,

IF

MONITOR/

CONTROLLER IS INSTALLED, ENSURE THE POWER

IS OFF. FAILURE TO DO SO COULD CAUSE DAMAGE

TO THE INSTRUMENT, AND COULD BE HARMFUL OR

FATAL

TO

PERSONNEL.

ONLY

QUALIFIED

PERSONNEL

SHOULD

INSTALL

OR

SERVICE

ELECTRICAL EQUIPMENT.

Physical

3RO Module only - A suitable location must be found that is both

convenient to the user and allow sufficient clearance for the 3

Range Switch Module to clear obstructions.

3RE - 3 Range Switch Module with enclosure, a cutout the size

of the enclosure is required. Install enclosure per Operation

Manual and skip to #4 below.

1. A 3/8” (9.5mm) hole must be drilled in the “user” panel.

This hole must be situated in a location that allows the

3 Range Switch Module to be mounted.

2. Install label, aligning on the drilled hole.

3. Mount the 3 Range Switch Module to the panel with

hardware supplied.

4. Remove the Range Module from the Main

Monitor/controller CB.

5. For Conductivity models, re-install Range Module on to

the 3 Range Switch Module in the LEFT location.

For Resistivity models, install replacement SPECIAL 3

Range-Range Module, part number RM3R, on to the

3 Range Switch Module in the LEFT location. In both

cases the arrow must be pointing toward the spot.

6. Attach cable harness from the 3 Range Switch Module

to the main Monitor/controller CB in the place of the

Range Module removed in step 4, as shown in

figure III.G.1.

7. Attach the separate 2 wire cable harness from the 3

Range Switch Module to the 2 pin male connector S3

located to the left of the sensor input connectors on

the main CB, as shown in figure III.G.1. NOT

INSTALLED ON MONITOR ONLY MODELS.

3. ALARM/CONTROL CONFIGURATION

As shipped from the factory, the 3R Module is configured so that

only the Primary or First switch position X1 is set to alarm/

control, as shown in figure III.G.2. If your particular applica-

tion requirements are different, the following will help you to

reconfigure the alarm/control function to another range or add a

controlling range. Pick the appropriate scenario below.

NOTE: the setpoints will not change nor are they separately

adjustable for each range.

If adding an alarm/control range.

1. Locate the 8 position select switch on the 3SR Module.

2. Set appropriate select switch UP or ON, i.e. to add

X100, set switch marked X100 UP or ON, see

figure III.G.3.

If changing alarm/control from one range to another, i.e. X1 to

X10 Range.

1. Locate the 8 position select switch on the 3SR Module.

2. Set the X1 select switch in the DOWN or OFF position.

3. Set the X10 select switch in the UP or ON position, see

figure III.G.4.

To disable either COND/TDS or Resistivity alarm/control func-

tion completely.

1. Locate the 8 position select switch on the 3SR Module.

2. Set all X prefixed type select switches in the DOWN or

OFF position, see figure III.G.5.

To reconfigure Conductivity to Resistivity or vise versa.

If you have a Resistivity Monitor/controller and ordered a

!

CAUTION - READ FOLLOWING CAREFULLY