Dvm self - powered nc remote - powered – Myron L 729II User Manual

Page 29

Advertising
background image

25

5. If not, adjust the CAL control marked “4mA” until

the DVM indicates 4mA, see figure III.B.3.

6. Press the Full Scale Test Switch, the DVM should

indicate 20 milliamps.

7. If not, adjust the CAL control marked “20mA” until the

DVM indicates 20mA. See figure III.B.3.

8. Calibration is complete

9. Turn power OFF.

10. FOR RESISTIVITY: Remove jumper from Black(BK)

and White(WT) connectors. For Conductivity, go to

step 11.

11. Reconnect sensor wires to sensor terminal block as

shown in figure III.B.4. (*For Resistivity models,

remove the jumper from the black and white

terminals and install the black and white sensor

wires.)

12. Carefully reinstall the front panel, bottom first, ensure no

wires have been pinched between enclosure and front

panel.

13. Reinstall the four (4) screws and tighten.

14. To operate, turn power ON.

MYRON L

COMPANY

Figure III.B.4.

SP1 HYS1

FUSE*

115/

230

}

DIS

3S

INC

SPC

0-10VDC

MAIN CIRCUIT BOARD ASSEMBLY

751 756

752 757

753 758

754 759

4-20

CONNECTOR

CONNECT 4AO

HARNESS HERE

FS SW

FULL SCALE

TEST SWITCH

PWR C GD NC NO CM

BK WT RD GN NU R- R+

SENSOR LEADS

TRANSFORMER

UP

CAL

CHS

GND

-121

2000µS

PA

RA

DEC

4-20

REMOVE T

O INST

ALL

SECOND RELA

Y

JUMPER

Figure III.B.5.

(+)

(-)

+

SO PO PI

SO PO PI

-

(+)

SIGNAL OUT

POWER OUT

POWER IN

+

DVM

Self - Powered

NC

Remote - Powered

(+)

(-)

+

-

(+)

SIGNAL OUT

POWER OUT

POWER IN

+

DVM

Remote

Power

Supply

NC

+

-

Advertising
This manual is related to the following products: