The following instructions describe the steps for converting the

Monitor/controller. Refer to figure III.A.4 for the location of the

SPC(s) on your Monitor/controller. If unnecessary, skip to Set

Point Adjustment.

1. Ensure power is OFF.

2. Locate the SPC jumpers for the alarm to be configured,

see figure III.A.4.

3. Make a note of the current orientation of the SPC

jumpers.

4. Remove both jumpers. This is easily done by hand.

Take care not to crush the jumpers if using pliers.

5. Rotate the jumpers 1/4 turn (90°) and reinstall them on

their posts.

6. Continue with Set Point adjustment.

b. Set Point Adjustment

Set point #1 (LOW) must be adjusted BEFORE adjusting set

point #2 (HIGH).

1. Turn power ON.

2. While depressing the lower “SET POINT” switch, turn

the Set Point #1 adjustment, figure III.A.4 until the

desired set point value is indicated on the display.

3. Repeat for Set Point #2 by depressing the upper “SET

POINT” switch and adjusting the SP2 on the SC/SCO

Module as shown in figure III.A.4.

4. Turn power OFF.

5. Continue or reinstall the front panel and tightly secure

both retaining screws, see REASSEMBLY below.

MYRON L

COMPANY

SOLID STATE (24VDC 30mA) Output

PA™ PIEZO ELECTRIC ALARM OR

RA™ REMOTE ALARM OR

CUSTOMER CONNECTION

SET POINT #1

CONVERSION

SET POINT #2

CONVERSION

COM

RELAY #2

}

NO

NC

SET POINT #1 HYSTERESIS

RIGHT INCREASING

LEFT DECREASING

SET POINT #2 HYSTERESIS

RIGHT INCREASING

LEFT DECREASING

SOLID STATE OUTPUT (24VDC 30mA)

PIEZO ELECTRIC ALARM - PA™ OR

REMOTE ALARM - RA™ OR

CUSTOMER CONNECTION

SET POINT #2 (high)

ADJUST

Figure III.A.4

CHS

GND

REMOVE T

O INST

ALL

SECOND RELA

Y

HYS1

SP1

FS SW

FUSE*

115/

230

-121

2000µS

UP

}

SC/SCO

OPTIONAL

SECOND

ALARM/

CONTROL

MODULE

#2 RELAY

LIGHTS &

SWITCH

CONNECTOR

TRANSFORMER

RUBBER TAPE

(DO NOT

REMOVE)

RELAY #1

}

COM

{

POWER

AC LINE/ +DC

AC NEUTRAL/ -DC

GROUND

NC

NO

PWR C GD NC NO CM

BK WT RD GN NU R- R+

CM NO NC

RA

SPC

INC

DIS

CAL

INC

3S

PA

FULL SCALE

PUSH TO TEST

751 756

752 757

753 758

754 759

INC

DEC

SPC

SET POINT #1 (low)

ADJUST