28

5. Set front panel down or carefully allow to hang from the

harness. Do not drop as the harness connector will pull

out allowing the front panel to fall.

6. While holding the display and the plastic display

retainer plate, carefully remove the display harness

connector. Do not drop the display.

7. Remove and discard the plastic display retainer plate.

8. Insert display connector pins into the Temperature

Module female connector. See figure III.C.1.

9. While holding the front panel, align the display to the

opening and at the same time, align the Temperature

Module mounting holes to the front panel.

10. Reconnect display harness with leads down as shown

in figures III.C.1 & III.C.2.

11. Reinstall the four (4) screws and tighten.

Display Select Switch installation

1. Using a small sharp knife or 1/4” (6.35mm) drill,

carefully cut open the hole on the right side of the front

panel, when viewed from the back, just below the Sensor

connectors on the Temperature module. See

figure III.C.2.

2. Install push-button display switch into this hole and

tighten bezel. See figures III.C.2 & III.C.4.

3. Install push-button/cap on switch.

WARNING: There are two (2) O-rings installed on the switch,

one (1) on the threaded shank and the other is under the push

button. Both of these must be re-installed to maintain IP64/

NEMA 3 ratings. See figure III.C.3.

4. Place “TEMPERATURE” label next to the switch. See

figure III.C.4.

Electrical

1. Connect the display switch harness to the Temperature

Module as shown in figure III.C.1.

2. Connect the Temperature Sensor leads to the

Temperature Module as labeled in figure III.C.1.

CAUTION: The sensor input connectors require only a small

screwdriver or a pen to push on the release levers. The release

levers may be broken or damaged if not pushed straight toward

the CB. DO NOT push the release levers sideways. Follow the

color code as labeled.

3. Connect 0-5VDC output, if desired.

4. To test, turn power ON.

5. Press “TEMPERATURE” front panel switch, display will

show the temperature of the sensor.

6. Turn power OFF.

7. Continue or reinstall the front panel and tightly secure

both retaining screws, see REASSEMBLY below.

3. RECALIBRATION

Electronic Calibration Only. For System Calibration, see below.

The Temperature Module was calibrated at the factory, however,

if you wish to check the calibration the following procedure will

help you to accomplish this task. Exercise caution while per-

forming this procedure.

Requirements:

TP Calibration Module (TPC) or

Two .1% precision resistors -

ZERO - 1000.0Ω (0°C = 0.0V)

SPAN - 1758.56Ω (200°C = 5.0V)

One (1) jumper

A DVM set to DC Volts, a tweaker or small standard slot

screwdriver.

FRONT PANEL

BEZEL

PUSH BUTTON

O-RING

O-RING

SWITCH

SWITCH and O-RING ASSEMBLY

Figure III.C.3

Figure III.C.4

FRONT PANEL

TEMPERATURE

Label Location

750II

HIGH

LOW

SET POINT

MICROSIEMENS / CM

TEMPERATURE

TEMPERATURE

Switch Location

Figure III.C.2

FRONT PANEL

Rear View

Cut front label

out for Display

Select Switch here

FS

ZERO

DIS

SP2

HYS2

DISPLAY CONNECTION

DISPLAY SWITCH

(UPPER FRONT PANEL)

DISPLAY

TEMPERATURE

SENSOR INPUT

0-5VDC

OUTPUT

DISPLAY

TEMP

CONTROL

HARNESS

TEMP

SENSOR

(-)

(+)

BLU

+

0-5VD

INC

INC

DEC DEC

BRN

BR

B

L

O

R

-

+

-

SPC

SELECT

TP/TPO - TEMPERATURE MODULE