Main cb assembly 15 – Myron L 729II User Manual

Page 19

Advertising
background image

F. 0-10 VDC RECORDER OUTPUT

The 0-10 VDC output is designed to give the user the capability

of sending a signal to a remote meter, recorder, PLC or SCADA

system.

1. CONNECTION

1. Place the user supplied interface cable and watertight

cable restraint into the enclosure’s appropriate access

hole. Skip for OEM.

2. Connect the Recorder’s plus (+) and minus (-) terminal

wires to the Recorder output connectors. (See figure

II.E.1.)

3. Refer to Section V.C.1.b for the procedures to calibrate

the 0-10 VDC voltage output.

2. VOLTAGE DIVIDER

A voltage divider gives the user the ability to scale or tailor the

output to a particular need or requirement due to the input of

another device, i.e. the output of the Main CB is 0-10V while the

input requirement of a particular recording device is 0-5V.

a. INSTALLATION

Briefly—

Two resistors are installed across the 0-10V output.

The output is recalibrated to required voltage.

Requirements

Select two (2) resistors as listed;

For 0-5V Output both “A & B” are 2K Resistors.

For 0-1V Output “A” is a 9K resistor and “B” is a 1K resistor.

WARNING: BEFORE STARTING, IF MONITOR/

CONTROLLER IS INSTALLED, ENSURE THE POWER IS

OFF. FAILURE TO DO SO COULD CAUSE DAMAGE TO

THE INSTRUMENT, AND COULD BE HARMFUL OR FATAL

TO PERSONNEL. ONLY QUALIFIED PERSONNEL SHOULD

INSTALL ELECTRICAL EQUIPMENT.

Physical

If the front panel has all ready been removed from the enclosure

skip to #3.

1. Using a standard slot screwdriver remove the four (4)

screws on the front panel.

2. Carefully wiggle the front panel to loosen and pull

gently toward you. Do not pull more than about

8 inches/20CM or you could damage the wiring

harness.

3. Turn the front panel around so that the back side is

facing you and set aside for now.

4. Solder two selected resistors together as shown in

figure II.F.1.

5. Attach leads to recording device as shown in figure

II.F.1.

6. Attach resistors to 0-10V Output as shown in figure

II.F.2.

Ensure resistors and leads DO NOT short to each

other or to any part of the CB assembly.

7. Recalibration is required, see Calibration Procedures,

section V.C.

Reassembly

1. Carefully reinstall the front panel, bottom first, ensure

no wires have been pinched.

2. Reinstall the four (4) screws and tighten.

3. To operate, turn power ON.

0-10V Output

+

-

A

B

+

-

Recording

Device

FS SW

BK WT RD GN NU R- R+

3S

SENSOR LEADS

CAL

0-10VDC

OUTPUT

NEW OUTPUT

AS SELECTED

+

-

Main CB Assembly

15

Figure II.F.1

Figure II.F.2

!

CAUTION - READ FOLLOWING CAREFULLY

Advertising
This manual is related to the following products: