Myron L 729II User Manual

Page 24

Advertising
background image

20

Front Panel

WARNING: There are two (2) O-rings installed on the switches,

one (1) on the shank and the other is under the push button.

Both of these O-rings must be reinstalled to maintain IP65/

NEMA 4X ratings. See figure III.A.3.

1-4 assumes this is a conversion from a single alarm/control.

Skip to #5 if new installation or assembly.

1. Carefully remove the RED LED from the front panel

by pulling lightly on the wires. It may be necessary to

use a small flat screwdriver to slightly spread the

retaining ring.

2. Remove the set point switch from the front panel by

rotating the round bezel nut CCW on the front panel,

see figure III.A.4.

3. Remove the LOWER front label.

4. Clean off remaining adhesive. Use alcohol, if possible.

5. Install DUAL alarm LOWER label, P# L2ALARM.

6. Install the SCO Second Alarm/Control RED LED in the

UPPER position as shown in figures III.A.5 & III.A.6.

7. Install the SCO Second Alarm/Control Set Point Switch

in the UPPER center position.

8. Re-install #1 (original) Set Point Switch in the LOWER

center position.

9. Re-install the #1 (original) RED LED in the LOWER

position.

10. The GREEN LED stays in the same location.

a. Set Point Conversion (SPC) /

Reversing Set Point

The alarm/control circuit(s) on all 750 Series II Conductivity/

TDS Monitor/controllers are configured to trigger the alarm relay

as the Conductivity/TDS reading increases.

NOTE: On Resistivity Monitor/controllers the reverse is true, as

the reading decreases the alarm/control will trigger/sound.

If the user’s application requires it, the alarm/control circuit

may be easily reconfigured to trigger the alarm/control relay as

the Conductivity (or ppm) reading decreases, or increases for

Resistivity. Refer to figure III.A.4 for the locations of the SPC

jumpers referred to in this section.

BK WT RD GN NU R- R+

RELAY #1

}

COM

{

POWER

AC LINE/DC+

AC NEUTRAL/DC-

GROUND

SET POINT #1

CONVERSION

CHS

GND

HYS1

SP1

FS SW

FUSE*

115/

230

NC

NO

-121

2000µS

UP

}

DIS

CAL

3S

PA

RA

REMOVE JUMPER TO

INSTALL SC/SCO

SECOND ALARM/CONTROL

MODULE

SET POINT #1 HYSTERESIS

RIGHT INCREASING /

LEFT DECREASING

FULL SCALE

PUSH TO TEST

SET POINT #1 (low)

ADJUST

INC

DEC

SPC

751 756

752 757

753 758

754 759

SOLID STATE OUTPUT

(24VDC 30mA)

PIEZO ELECTRIC ALARM - PA™

OR

REMOTE ALARM - RA™ OR

CUSTOMER CONNECTION

PWR C GD NC NO CM

TRANSFORMER

REMOVE

TO INST

ALL

SECOND RELA

Y

Figure III.A.2

SWITCH

FRONT PANEL

BEZEL

NUT

PUSH

BUTTON

O-RING

O-RING

Switch / O-ring Assembly

Figure III.A.3