Design and function – WIKA LSO.25 User Manual

Page 13

Advertising
background image

WIKA operating instructions models LSO.06, LSO.25

13

13334603.01 09/2010 GB/D/F/E

4. Design and function

GB

4. Design and function

4.1 Description

The transducer is used for measuring liquid level limits.

The conical tip of the sensor ensures really precise on-off performance level sensing.

Its reaction is independent of various physical parameters of the liquid such as density,

dielectric constant, conductivity, colour or refractive index. This allows the use for safe

level switching as well as for precise level control applications. If there is any foam you

even can decide whether to use the sensor for foam level or liquid level control, suppress-

ing the influence of foam on switching.

If the sensor is supplied with a U-tip, it is able to sense changes in refractive index. So it

can be used for example as an interface level switch for liquids.

Application range

Principle of operation

Separation layer, U-tip

e.g. oil

e.g. water

Principle of operation

Level, conical tip

Gas
Liquid

Working pressure in bar

Working pressure in bar

Medium temper

atur

e in °C

Medium temper

atur

e in °C

Standard pressure version

without heatsink

threaded connection

High pressure

version with

heatsink

threaded

connection NPT ½

High pressure

version with

heatsink

threaded

connection G ½ A

All typs of transducers need a switching amplifiers model LSO.25 for power supply and

switch state signalling.

Advertising