Maintenance, repair and cleaning – WIKA LSO.25 User Manual

Page 31

Advertising
background image

WIKA operating instructions models LSO.06, LSO.25

31

13334603.01 09/2010 GB/D/F/E

GB

Replacing fuse of switching amplifier

Switch net supply off

Taking fuse out

Use a new fuse only with the value printed on the circuit board

Supply voltage

Model LSO.06

AC 230 V

T 50 mA

AC 115/120 V

T 100 mA

AC 24 V

T 400 mA

DC 24 V with potential separation

see marking on circuit board

DC 24 V without potential separation

T 400 mA

Switch net supply on

At spare part orders indicate absolutely exact type, series no. and

commission no. the manufacturer.

9.2 Cleaning
If the glass tip becomes soiled, clean it carefully.

CAUTION!

Before cleaning, correctly disconnect the instrument from the pressure

supply, switch it off and disconnect it from the mains.

Clean the instrument with a moist cloth.

Electrical connections must not come into contact with moisture.

Wash or clean the dismounted instrument before returning it, in order to

protect staff and the environment from exposure to residual media.

Residual media in dismounted instruments can result in a risk to persons,

the environment and equipment. Take sufficient precautionary measures.

For information on returning the instrument see chapter "12.2 Return".

9. Maintenance, repair and cleaning