Normal operating conditions, Performance, Front panel controls – Bacharach Portable Combustion Analyzer 24-9351 User Manual

Page 10

Advertising
background image

PCA

Instruction 24-9351

Normal Operating Conditions:

Temperature:
Analyzer ................... 0 to 40 ºC (32 to 104 ºF)
Probe ......................... 800 ºC (1472 ºF) Max.

Humidity:
Analyzer ................... 15 to 90% Relative Humidity, Non-Condensing

Air Pressure:
Analyzer ................... Atmospheric
Probe ......................... 25 mb (10" H

2

O) draft max at probe tip

Performance:

Accuracy:
Oxygen* .................... ±0.3% O

2

Carbon Monoxide ..... ±5% of reading or ±10 ppm, whichever is greater

between 0 – 2000 ppm, and ±10% of reading
between 2001 – 4000 ppm.

Nitric Oxide .............. ±5% of reading or ±5 ppm, whichever is greater
Flue Gas Temp. ........ ±2 ºC between 0 and 124 ºC

(±4 ºF between 32 and 255 ºF)

±3 ºC between 125 and 249º C

(±6 ºF between 256 and 480 ºF)

±4 °C between 250 and 400 °C

(±8 ºF between 481 and 752 ºF)

Ambient Temp. ........ ±1 °C between 0 and 100° C

(±2 ºF between 32 and 212 ºF)

Pressure ................... ±2% of reading or ±0.05 mb (±0.02 inches of

Water Column), whichever is greater

System Flow Rate:
With probe ................ 200 cc/min minimum

Front Panel Controls:

Seven embossed push-button switches with tactile feedback (refer to
Section 4.1)

* Accuracy referenced in practical flue gas concentrations

(mixtures of O

2,

CO

2

and N

2

)

2-2

Technical Characteristics