Bacharach Portable Combustion Analyzer 24-9351 User Manual

Page 76

Advertising
background image

PCA

Instruction 24-9351

NX-SENSOR ERROR code ap-
pears in the Sensor Status Screen.

CO-SENSOR ERROR code ap-
pears in the Sensor Status Screen.

DRAFT-SENSOR ERROR code ap-
pears in the Sensor Status Screen.

TA-SENSOR OR TL-SENSOR
ERROR
code appears in the Sen-
sor Status Screen.

****” appears in one or more value
fields.

7-4

TABLE 7-1. TROUBLESHOOTING GUIDE (Cont.)

Fault

Probable Cause & Remedy

Troubleshooting

a. Calibration was attempted while

sampling combustion gases.

b. Nitric Oxide sensor is expired.

Replace sensor per Section 6.7.

c. Bias battery on the Nitric Oxide

circuit board is dead. Replace bat-
tery per Section 6.7.2.

a. Calibration was attempted while

sampling combustion gases.

b. Carbon Monoxide sensor is ex-

pired. Replace sensor per Sec-
tion 6.8.

a. Sensor is exposed to pressure out-

side of its detectable range.

b. Sensor defective. Return analyzer

to Bacharach for repair.

a. Calibration was attempted while

sampling combustion gases.

b. Thermocouple defective. Re-

place probe assembly.

c. Thermocouple not connected.

Connect thermocouple to ana-
lyzer per Section 3.3.

d. Instrument exposed to temp-

erature outside it’s allowable
operating temperature.

The field’s associated sensor is not
installed.