4 differential pressure measurement, 4 differential pressure measurement -7 – Bacharach Portable Combustion Analyzer 24-9351 User Manual

Page 23

Advertising
background image

Instruction 24-9351

PCA

4-7

4.4 Differential Pressure Measurement

The difference in pressure (

∆P) between two areas can be measured by

using the PCA’s two pressure ports and DRAFT Screen. By using Pres-
sure Port 2 (–) as the reference, the pressure applied to Port 1 (+) will be
displayed on the DRAFT Screen as the differential pressure between the
two ports.

1. Turn on the analyzer by pressing the

key; wait for the warmup cycle

to complete; then press the

MENU

key until the first DRAFT Screen is

displayed (refer to Section 4.9). If a TA-SENSOR ERROR is displayed
because the probe’s thermocouple is not plugged into the analyzer, then
press the

key to acknowledge the error before pressing the

MENU

key.

2. While the first DRAFT Screen is displayed, remove any hoses con-

nected to Pressure Ports 1 and 2; then press the

key to zero these

ports at atmospheric pressure.

3. Connect two sampling hoses to

Pressure Ports 1 and 2
(Figure 4-3). Then place the open
end of each hose into the areas
being measured.

4. The differential pressure between the two areas is now displayed on

the third DRAFT Screen. If the pressure at Port 1 is higher than
Port 2, then the pressure difference will be positive. But if the pres-
sure at Port #1 is lower, then the pressure difference will be negative.
The reading shown in this example indicates that the pressure at
Port 1 is 2.25 mb lower than the pressure at Port 2.

Operation

Figure 4-3. Differential Pressure Hose Connections

DRAFT

DRAFT – 2.25 MB

HOT SPOT ---- °C P

«S

Pressure Port 1(+)

Pressure Port 2(–)
Reference

Sampling Hoses with quick-
connect fittings are available
as an optional accessory.
Refer to Section 8.2.