7.3 Troubleshooting Guide

The following table lists the causes and remedies for most of the problems
that may arise with the analyzer. For help with any problem not dis-
cussed here, contact your nearest Bacharach Service Center.

TABLE 7-1. TROUBLESHOOTING GUIDE

Fault

Probable Cause & Remedy

Analyzer completely nonfunctional;
won’t turn on when the

key is

pressed.

Display Screen is blank when ana-
lyzer is turned on, but pump runs
during warmup cycle.

LOW BATTERY message appears
at bottom of display.

O2-SENSOR ERROR code ap-
pears in the Sensor Status Screen.

a. Batteries dead. Replace batteries

per Section 3.2.

b. Loose battery connector. Disas-

semble analyzer and ensure that
the battery connector is attached
to the printed circuit board.

c. Microprocessor needs to be reset.

Press RESET button (Fig. 4-7).

d. Optional Power Supply defective.

Replace Power Supply.

e. Analyzer defective. Return to

Bacharach for repair.

a. Microprocessor needs to be reset.

Press RESET button (Fig. 4-7).

b. Analyzer defective. Return to

Bacharach for repair.

Batteries close to being discharged.
Analyzer will run for several min-
utes before the instrument shuts off.
Replace batteries per Section 3.2.1.

a. Calibration was attempted while

sampling combustion gases.

b. O

2

sensor is either expired, not

wired correctly, or not connected to
the circuit board. Replace or check
wiring of sensor per Section 6.6.

7-3

Troubleshooting