Instruction 24-9351

PCA

5.5 Calibrate TA-Span

Material Required:

Thermocouple Simulator (K-type)

Range: 0 to 300°C
Accuracy: ±0.3°C

Procedure:

1. With the analyzer turned off, first plug the simulator's K-type connec-

tor into the T-STACK jack (Figure 3-2); then turn on the analyzer and
wait for its warmup cycle to complete.

2. Enter the Calibration Menu Screen per Section 5.3; then choose

TA-SPAN to span the analyzer’s stack temperature channel.

3. Set the simulator to 300° C (572° F).*

4. Wait until the MEASURED reading on the screen stabilizes. Then use

the

st

and

keys to enter an APPLIED value that equals 300 °C

(572 °F).

Typical Calibrate TA-Span Screen During Calibration Procedure:

CALIBRATE TA-SPAN

MEASURED 295.0°C

APPLIED 0300.0°C

4. Press the

key to calibrate the analyzer’s MEASURED value to

that of the APPLIED value. At this time the Calibrate Menu Screen is
redisplayed.

* The calibration range for this screen is 270 – 330 °C (518 – 626 °F).

Any attempt to calibrate outside this range will cause the analyzer to
display an error message.