9 calibrate co, 9 calibrate co -9 – Bacharach Portable Combustion Analyzer 24-9351 User Manual

Page 57

Advertising
background image

Instruction 24-9351

PCA

5.9 Calibrate CO

(For PCA Models 20, 25, 30, 35, 50, 55, 60 & 65)

Material Required:

Calibration Gas Fixture (Section 5.2)
Gas Cylinder, 500 ppm CO in air (Refer to Section 8.2)
Gas Cylinder, CO (1000 ppm) and H

2

(1000 ppm) in Nitrogen

(Refer to Section 8.2)

Procedure:

1. Enter the Calibrate Menu Screen per Section 5.3. Then choose to

calibrate the CO sensor.

2. At the conclusion of Step 1 the pump should start running.

3. Using the Gas Fixture shown in Figure 5-1, attach the CO calibration-

gas cylinder to the analyzer’s GAS inlet.

4. Adjust the regulator of the calibration fixture for a flowmeter indica-

tion of approximately 2 SCFH.

5. Wait until the MEASURED reading on the screen stabilizes (approxi-

mately 3 minutes). Then use the

st

and

keys to enter an

APPLIED value* that equals the concentration which is stamped on
the CO calibration-gas cylinder.

Typical CALIBRATE CO Screen During Calibration Procedure,
Using 500 ppm CO Calibration Gas:

CALIBRATE CO

MEASURED 492 PPM

APPLIED 0500 PPM

6. Press the

key to calibrate the analyzer’s MEASURED value to

that of the APPLIED value. At this time the Test Gas CO/H2 Screen
is displayed.

* The calibration range for this screen is 250 – 1500 ppm. Any attempt

to calibrate outside this range will cause the analyzer to display an
error message.