PCA

Instruction 24-9351

5.6 Calibrate TL-Zero

Material Required:

Thermocouple Simulator (K-type)

Range: 0 to 300°C
Accuracy: ±0.3°C

Procedure:

1. With the analyzer turned off, first plug the simulator’s K-type connec-

tor into the T-AIR jack (Figure 3-2); then turn on the analyzer and
wait for its warmup cycle to complete.

2. Enter the Calibration Menu Screen per Section 5.3; then choose

TL-ZERO to zero the analyzer’s room-air/primary-air temperature
channel.

3. Set the simulator to 0 °C (32 °F).*

4. Wait until the MEASURED reading on the screen stabilizes. Then use

the

st

and

keys to enter an APPLIED value that equals 0 °C

(32 °F).

Typical Calibrate TL-Zero Screen During Calibration Procedure:

CALIBRATE TL-ZERO

MEASURED 3.0°C

APPLIED 0000.0°C

4. Press the

key to calibrate the analyzer’s MEASURED value to

that of the APPLIED value. At this time the Calibrate Menu Screen is
redisplayed.

* The calibration range for this screen is 0 – 5 °C (32 – 41 °F). Any

attempt to calibrate outside this range will cause the analyzer to
display an error message.

Calibration

5-6