PCA

Instruction 24-9351

7. Turn off the flow of CO calibration-gas; then remove the calibration-

gas cylinder from the calibration fixture.

8. Attach a CO/H

2

calibration-gas cylinder to the calibration fixture;

then adjust the regulator of the calibration fixture for a flowmeter
reading of approximately 2 SCFH.

9. Use the

st

and

keys to enter a CO-VALUE that is the same as

the CO concentration which is stamped on the CO/H

2

calibration-gas

cylinder.

Typical Test Gas CO/H2 Screen During Calibration Procedure:

TEST GAS CO/H2

-------------------

CO-VALUE XXXX

ENTER CO-VALUE

10. Press the

key to save the CO-VALUE and display the Calibrate H2

Screen.

11. After calibration gas has been applied for approximately 3 minutes (to

allow for stabilization), use the

st

and

keys to enter an APPLIED

value* that equals the H

2

concentration stamped on the CO/H

2

calibration-gas cylinder.

Typical CALIBRATE H2 Screen During Calibration Procedure:

CALIBRATE H2

MEASURED 1050 PPM

APPLIED 1000 PPM

12. Press the

key to calibrate the analyzer’s MEASURED value to

that of the APPLIED value. At this time the Calibrate Menu Screen is
redisplayed.

* The calibration range for this screen is 500–1500 ppm, any attempt to

calibrate outside this range will cause the unit to display an error
message.