9 draft screens, 10 memory directory screen -13 – Bacharach Portable Combustion Analyzer 24-9351 User Manual

Page 28

Advertising
background image

PCA

Instruction 24-9351

4-12

Operation

4.9 Draft Screens

The first Draft Screen is displayed by
repeatedly pressing the

MENU

key

from the Combustion Test Screen.

To measure draft, first zero the
analyzer’s pressure sensor to atmo-
spheric pressure by disconnecting the
draft hose from the bottom of the
instrument, and then pressing the

key. Reconnect the draft hose after

the second Draft Screen appears
(shown for 3 seconds). The third
screen shows the current values of
draft and stack temperature as
measured by the analyzer.

When using the analyzer to make a
differential pressure measurement
(Section 4.4), the differential pressure
value will be displayed on the third
Draft Screen.

To save (

S) or print (P) the screen

data, first use the

st

keys to move

the cursor (

z

) in front of the desired

function, and then press the

key.

Front Panel Key Functions:

– Move cursor (

z

) up

– Move cursor (

z

) down

– Save or Print screen data

– Go to Combustion Test Screen

– Go to Memory Directory Screen

DRAFT

DISCONNECT DRAFT

HOSE

PRESS ±

DRAFT

RECONNECT DRAFT

HOSE

DRAFT

DRAFT – 0.25 XX

HOT SPOT 190 °C

P

«

S

Where: xx = Unit of measure.
Default is millibars (MB). See
Optional Draft SETUP Screen
(Section 4.14) for other choices.