Sterile container system

Aesculap Sterile Technology

6. Troubleshooting list

Malfunction

Cause

Remedy

Excessive amounts of condensate inside the
sterile container

Temperature of sterile materials too low prior
to sterilization

Allow sterile materials to come to room
temperature (appr. 20 °C) before sterilization

Textiles too damp

Sterilize dry textiles only

Sterile container too heavy

Normal container
with instruments: max. load 10 kg
with textiles: max. load 8 kg

Half container with instruments:
max. load 5 kg

3/4-container with instruments: max. load 7 kg

Sterile materials improperly positioned

Lay hollow components, bowls, plates etc. face
down and on a slant

Arrange textiles in loose vertical piles, do not
press them together

Sterile container incorrectly positioned in
sterilizer

Always place heavy sterile containers at the
bottom

Sterile containers processed immediately
after sterilization

Allow sterile containers to cool down to room
temperature prior to processing

Sterile containers improperly positioned
during cooling phase

Do not store sterile containers on a floor or in a
drafty place
Store sterile containers in air-conditioned areas
according to DIN 58953-9

Sterilizer properties do not meet DIN EN 285
standards

Have sterilizer serviced regularly
Check drying vacuum
Check drying time

Check steam quality and upgrade if necessary

Empty-cycle and vacuum test not run daily
before sterilization begins

Run empty-cycle and vacuum test daily before
beginning sterilization

Unsuitable sterilizer cycle selected

Select a program according to the load

Sterilizer door left open too long, sterilizer
cooled down

Load and unload sterilizer quickly

Condensate in groove of (inner) lid

Multiple sterile containers or heavy sterile
materials stacked

Subdivide sterilizer with shelving
Use outer lid

Indicator seal color unchanged

Sterilization performed incorrectly
Sterilizer defective

Have sterilizer repaired by manufacturer

Indicator seals stored improperly

Observe storage conditions as specified on the
packaging of the indicator seals