T3000 – INFICON T3000 Thermal Power Supply User Manual

Page 10

Advertising
background image

instruments

Thermal Power Supply

T3000

Section 1

Page 1 - 3

Introduction – Description

of Equipment

1.4 Output

Characteristics


The purpose of this subsection is to provide guidelines for choosing between series or
parallel output connections based on the load requirements of the thermal source being
energized. Output voltage and current profiles vary in proportion to load resistance, up to
maximum limits determined by the output circuit configuration. Figure 1-1 provides a
guideline for choosing the appropriate output hookup configuration, based on the T3000
maximum output profile and the load requirements of the thermal source.

Maximum Load Voltage-Current Profile

0

2

4

6

8

10

12

14

16

0

100

200

300

400

500

600

700

Load Current - Amps RMS

Load V

o

ltage - V

o

lts RMS

Max Output: Parallel Configuration

Max Output: Series Configuration

Series Only Region

Parallel Region Only

Either Series or Parallel Region

Vline = 220V rms

Figure 1-1

Maximum Voltage-Current Profile for Series and Parallel Hookup


Figure 1-1 is marked with regions which define when the series or parallel output
configuration is appropriate. The maximum voltage-current rating of any given thermal
source may be located as a point on the graph. The appropriate configuration(s) is
determined by the location of this point.

For example, consider a commercially available thermal source rated at 521 ARMS and a
corresponding voltage drop of 3.89 VRMS. This is a 2kW thermal source and falls in the
area of the graph noted as “Parallel Only”.

By way of contrast, a thermal source rated at 742 Watts, which is specified by a voltage
drop of 8.73 VRMS at 85 ARMS and falls in the “Series Only” region of the graph.

Advertising