Section 6

Table 6-2

Inspection and Corrective Action

ITEM

INSPECT FOR

CORRECTIVE ACTION

Connector plugs and buss
bars

Loose, bent or corroded
contacts and buss bars.
Damage or improper seating
in mating connector.

Clean contacts and buss
bars with solvent moistened
cloth, soft bristle brush, small
vacuum or low compressed
air.

Chassis, fans & extruded
heatsinks

Dirt and corrosion

Replace damaged or
corroded connectors.

External electrical wiring

Broken, burned or pinched
wire. Frayed, worn or
missing insulation

Clean with cloth moistened
with soapy water.

External solder connections

Corrosion, loose, cracked, or
dirty connections

Repair or replace defective
wires.

Dirt and moisture buildup

Short circuits, arcing,
corrosion, overheating

Clean and resolder
connections.

T3000 enclosure

Dirt and corrosion

Clean as required with cloth
moistened with soapy water.