T3000 – INFICON T3000 Thermal Power Supply User Manual

Page 43

Advertising
background image

T3000

Thermal Power Supply

instruments

T3000 Configurations

Page 7 - 4

Section 7

7.4

T3000 Controlled by STC via RS3000


1) Verify that the AC input power is disabled.

2) Connect the AC-In power cable to the T3000 as shown in Figure 2.2-1.

3) Connect the output buss bars for the desired configuration (Series or Parallel), per
Figures 2.2-3 or 2.2-4.

4) Position the SER-PAR slide switch to coincide with the output buss bar configuration.
(Note: Switch positioned to the left is for the Series configuration while switch positioned to
the right is for the Parallel configuration,)

5) Connect the control cable from the RS3000 to J1 as shown in Figure 3.2-1.

6) Connect the cable from the STC output to JX of the RS3000.

7) Apply the AC power. The green LED should now “blink” indicating that AC power is
applied to the T3000.

8) Apply appropriate input parameters to the STC200 or STC2000, as described in their
respective manuals, and start the process. The green LED should now be on in a steady
state mode, indicating that the T3000 is ready to supply current to the load. (Note: Set STC
output for the 0 to +10 V range).