T3000 – INFICON T3000 Thermal Power Supply User Manual

Page 42

Advertising
background image

instruments

Thermal Power Supply

T3000

Section 7

Page 7 - 3

T3000 Configurations

7.3 T3000

Controlled by

STC


1) Verify that the AC input power is disabled.

2) Connect the AC-In power cable to the T3000 as shown in Figure 2.2-1.

3) Connect the output buss bars for the desired configuration, series or parallel, as per
Figures 2.2-3 or 2.2-4.

4) Connect the control cable to J1 as shown in Figure 3.2-1. (Note: A custom split cable is
required if voltages measured on pins 5 and 14 are used as monitor points.)

5) Apply the AC power. The green LED should now “blink” indicating that AC power is
applied to the T3000.

6) Apply appropriate input parameters to the STC200 or STC2000, as described in their
respective manuals, and start the process. The green LED should now be on in a steady
state mode, indicating that the T3000 is ready to supply current to the load. (Note: Set STC
output for the 0 to +10 V range).