Section 2

Page 2 - 1

Installation

Section 2

Installation

2.1 Inspection

Inspect the shipping carton for possible damage before unpacking the unit. Carefully
unpack the equipment. Save all packing materials until inspection is complete. Verify that
all items listed on the packing slips have been received. Visually inspect all exterior
surfaces for damage. External damage may be an indication of internal damage. If any
damage is evident, immediately contact the carrier that delivered the unit and submit a
damage report. Failure to do so could invalidate future claims.

2.2 Input/Output

Connectors


Table 2.2-1 lists all external connections for the T3000. An input and output connector
description is provided in Table 2.2-2. For permanently connected equipment, a readily
accessible safety switch shall be incorporated in the fixed AC wiring.







Table 2.2-1

T3000 External Connections

CONNECTOR FUNCTION

CONNECTS

TO

AC1 (See Figure 2.2-1)
AC2
Frame Ground

Section 1 (See Figure 2.2-2)
Positive Bus Bar
Negative Bus Bar



Section 2 (See Figure 2.2-2)
Positive Bus Bar
Negative Bus Bar

J1

Prime AC Power Input
Single Phase


Output Power





Output Power



Control Interface

180-264 VAC
47-63 Hertz
Power Source

User Load

(See Figure 2.2-3 for
series or parallel
connections)

User Load



See Table 3.2-1

NOTICE

For proper connection to the mains, a 30-Amp circuit
breaker or fuse is required in each AC line.