T3000 – INFICON T3000 Thermal Power Supply User Manual

Page 41

Advertising
background image

T3000

Thermal Power Supply

instruments

T3000 Configurations

Page 7 - 2

Section 7

7.2

T3000 Controlled by RS3000


1) Verify that the AC input power is disabled.

2) Connect the AC-In power cable to the T3000 as shown in Figure 2.2-1.

3) Connect the output buss bars for the desired configuration, series or parallel, as per
Figures 2.2-3 or 2.2-4.

4) Position the SER-PAR slide switch to coincide with the output buss bar configuration.
(Note: Switch positioned to the left is for the Series configuration while switch positioned to
the right is for the Parallel configuration,)

5) Connect the control cable to J1 as shown in Figure 3.2-1.

6) Verify that all of the interlocks to the RS300 are satisfied.

7) Apply the AC power. The green LED should now “blink” indicating that AC power is
applied to the T3000.

8) Apply appropriate input parameters to the RS3000 as described in Section 7 and
depress the Start control of the RS300. The green LED should now be on in a steady state
mode that indicates that the T3000 is ready to supply current to the load. (Note: The
RS3000 can be operated in the programmed or manual mode.) Refer to Section 7.X for
description of RS3000 displays.