2 electrical connection – INFICON SKY CDG025D-X3 0.1 ... 1000 Torr / mbar process gauge User Manual

Page 16

Advertising
background image

16

tina57e1 (2010-01)

3.2 Electrical

Connection

Make sure the vacuum connection is properly made

(

→ 13).

DANGER

The gauge may only be connected to power sup-
plies, instruments or control devices that conform
to the requirements of a grounded protective extra-
low voltage (SELV) and limited power source
(LPS), Class 2. The connection to the gauge has to
be fused

4)

.

Ground loops, differences of potential, or EMC problems

may affect the measurement signal. For optimum signal
quality, please do observe the following notes:

• Connect the cable shield to ground on one side via

the chassis ground. Do not connect the other side of
the shield.

• Connect the supply common with protective ground

directly at the power supply.

• Use differential measurement input (signal common

and supply common conducted separately).

• Potential difference between supply common and

housing ≤18 V (overvoltage protection).

4)

INFICON controllers fulfill this requirement.

Advertising