Off on, 1 adjusting the setpoints – INFICON SKY CDG025D-X3 0.1 ... 1000 Torr / mbar process gauge User Manual
Page 24
Advertising
24
tina57e1 (2010-01)
If the pressure is lower than the setpoint, the corresponding LED
is lit (<1> or <2>) and the corresponding relay (
→ 17) is
energized.
Measurement signal (Pressure p)
Time t
Me
asu
reme
nt value
Off
On
Hysteresis
(1% F.S.)
Off
Threshold value
Setpoint
4.3.1 Adjusting the Setpoints
The setpoints can be adjusted via
• the buttons on the gauge,
• the RS232C interface (→ [4]).
DANGER
DANGER: malfunction
If processes are controlled via the signal output,
keep in mind that by pushing the <SP> button the
measurement signal is suppressed and the cor-
responding threshold value is output instead. This
can cause malfunctions.
Push the <SP> button only if you are sure that no
damages can arise from a malfunction.
Advertising