INFICON SKY CDG025D-X3 0.1 ... 1000 Torr / mbar process gauge User Manual
Page 21
Advertising
tina57e1 (2010-01)
21
Briefly press the <ZERO> button with a pin (max.
ø1.1 mm). The zero adjustment runs automatically. The
<STATUS> LED flashes until the adjustment (duration
≤8 s) is completed.
Press the button briefly
max. ø1.1 mm
After zero adjustment the gauge automatically returns to
measurement mode.
The <RUN> LED flashes if
• the signal output is negative (< -20 mV) when the final pres-
sure has been attained
• the zero adjustment has failed.
4.2.2
<ZERO> Adjustment with Ramp Function
The ramp function allows to adjust the zero at a known reference
pressure within the measurement range of the gauge.