3 switching functions – INFICON SKY CDG025D-X3 0.1 ... 1000 Torr / mbar process gauge User Manual
Page 23
tina57e1 (2010-01)
23
Push the <ZERO> button again:
Fine adjustment
within 0...3 s:
the zero adjustment value changes
by one unit (push <ZERO> button in
intervals of 1 s)
Change of direction
within 3...5 s:
the zero adjustment changes its
direction (the flashing frequency of
the <RUN> LED changes briefly)
If the <ZERO> button is released for more than 5 s, the
gauge returns to the measurement mode.
The <RUN> LED flashes if the signal output is negative (< -
20 mV).
4.3 Switching
Functions
The two switching functions can be adjusted to any pressure
within the whole measurement range (
→ 12).
The current setpoint setting
• is output at the D-Sub connector instead of the measurement
signal (
→ 17) and can be measured with a voltmeter after
the <SP> button is pressed, or
• can be read/written via the RS232C interface.