2 zeroing the gauge – INFICON SKY CDG025D-X3 0.1 ... 1000 Torr / mbar process gauge User Manual

Page 19

Advertising
background image

tina57e1 (2010-01)

19

4.2

Zeroing the Gauge

The gauge is factory calibrated while "standing upright"
(

→ "Calibration Test Report").

We recommend performing a zero adjustment, when the

gauge is operated for the first time.

Due to long time operation or contamination, a zero drift could
occur and zero adjustment may become necessary.
For adjusting the zero, operate the gauge under the same con-
stant ambient conditions and in the same mounting orientation
as normally.
The output signal (measuring signal) is depending on the moun-
ting orientation. The signal difference between the vertical and
horizontal mounting orientation is:

F.S.

Δ

U / 90°

1000 Torr/mbar

≈2 mV

100 Torr/mbar

≈10 mV

10 Torr/mbar

≈50 mV

1 Torr/mbar

≈300 mV

0.1 Torr/mbar

≈1.8 V

If the gauge is operated via a controller, the zero of the

whole measuring system has to be adjusted on the con-
troller: first, adjust the zero of the gauge and then, the
zero of the controller.