Technical data – Leica Biosystems CM1850 UV User Manual
Page 12
12
Instructions for use V 1.5 – 10/2012
Operating temperature range (ambient temperature): 18 °C to 35 °C.
All specifications related to temperature are valid only up to an ambient temperature of 22 °C
and an air humidity lower than 60 %!
3.
Technical data
Type
CM1850 UV
CM1850 UV
CM1850 UV
CM1850 UV
CM1850 UV
Mark of conformity
-
CUL
-
-
VDE
Nominal voltage
100 V AC ±10%
120 V AC ±10%
220 V AC ±10%
230 V AC ±10% 240 V AC ±10%
Nominal frequency
50/60 Hz
60 Hz
60 Hz
50 Hz
50 Hz
Power input
1440 VA
1440 VA
1440 VA
1440 VA
1440 VA
Max. start-up current for 5 sec 45 A eff.
45 A eff.
45 A eff.
45 A eff.
45 A eff.
Protective class
I
I
I
I
I
Automatic mains fuse
T12A T1
T12A T1
T12A T1
T10A T1
T10A T1
Pollution degree
☛
2
2
2
2
2
Overvoltage inst. category
II
II
II
II
II
Heat emission (max.)
1600 J/s
1600 J/s
1600 J/s
1600 J/s
1600 J/s
Refrigeration system
50 Hz
60 Hz
Cryochamber
Temperature setting range
0 °C to -35 °C (+ 2 K/- 0 K)
0 °C to -35 °C (+2 K/- 0 K)
Defrosting
automatic hot gas defrosting
automatic hot gas defrosting
1 automatic defrost cycle/24 hrs,
1 automatic defrost cycle/24 hrs,
temperature controlled;
temperature controlled;
Refrigeration capacity
☛
690 W
690 W
Safety factor
3
3
Refrigerant
300g ±5g refrigerant R 404A *
300 g ±5g refrigerant R 404A*
Compressor oil
0.6 l EMKARATE RL-22S, ICI *
0.6 l EMKARATE RL-22S, ICI *
Quick-freeze shelf
Max. temperature
- 43 °C (+3 K/-2 K)
- 43 °C (+3 K/-2 K)
No. of quick-freeze stations
10
10
Defrosting
manual hot gas defrosting,
manual hot gas defrosting,
time controlled
time controlled
Peltier element (optional)
Lowest possible temperature
-60 °C (+5K)
-60 °C (+5K)
No. of quick-freeze stations
2
2
Defrosting
Together with quick-freeze shelf
Together with quick-freeze shelf
☛
according to CECOMAF Liquid temperature 45 °C, evaporation temperature -25 °C
☛
according to IEC-1010, UL 3101